Breath Testing in North Dakota DUI Cases
A breath-test printout looks simple because the complicated work happened before the number appeared. The officer selected the test, watched the subject…
North Dakota DUI breath test, Fargo Intoxilyzer 8000 defense, approved breath-test method, twenty-minute deprivation period, mouth alcohol, Difference Too Great, deficient sample, operator certification, instrument records, fair administration, NDDOT hearing, independent chemical test, and breath-test expert
Breath Testing in North Dakota DUI Cases
A breath-test printout looks simple because the complicated work happened before the number appeared. The officer selected the test, watched the subject, operated a particular instrument under a date-specific approved method, entered identifying information, collected breath samples, and relied on the machine’s diagnostic, air, and gas-standard checks. Every step leaves records that can be compared with video, testimony, and the governing method.
The defense question is not whether an Intoxilyzer can measure alcohol. It is whether this operator fairly administered this test on this device, whether the statutory foundation is complete, and whether the reported concentration answers the charge actually filed. The criminal prosecution and the NDDOT license case use related evidence but remain separate proceedings with different deadlines and burdens.
North Dakota law requires a proper sample and a fairly administered test
The statutory shortcut has conditions
Section 39-20-07 does not say every number printed by an instrument is admissible. Subsection 5 requires proof that the sample was properly obtained, the test was fairly administered, the method and device were approved, and the person administering the test was qualified. Subsection 6 directs the Crime Laboratory to approve satisfactory techniques, methods, and devices and to determine qualifications for the people who conduct chemical analyses.
Subsection 7 allows certified copies of posted records to establish prima facie that a device, operator, method, or checklist was approved. “Prima facie” means the documents can establish the point unless countered; it does not prevent scrutiny of identity, dates, completeness, deviations, or contradictory evidence. Match every certificate to the offense date, operator number, instrument serial number, location, method revision, and actual test record.
The effective method is the method in force on the test date
The Crime Laboratory publishes the approved Intoxilyzer 8000 methods. The current page includes the April 22, 2025 method with addendum and earlier versions. A later clarification should not be silently substituted for the document governing an earlier test. Obtain the correct version, its addenda, the operational checklist, and any contemporaneous laboratory memorandum.
A lawyer should compare the method line by line with the test record, checklist, and video. The important question is not merely whether a box was checked. It is what occurred, who observed it, when it occurred, whether the record accurately reflects it, and whether any departure could affect the scientific accuracy or reliability of the result.
The operator, instrument, and gas standard each have an identity
The printout should identify the North Dakota Model 8000 and serial number. The operator uses a unique certification number, and the standard-gas information includes lot, cylinder, and expiration data. The Attorney General publishes the device records, including inspections and ethanol-gas-standard documents organized by instrument.
Confirm that the operator was certified on the test date, the instrument appeared on the approved-device register, the device record corresponds to the printed serial number, and the gas cylinder was approved and unexpired. A nearby inspection date is not enough if the test involved another device, cylinder, or operator. Conversely, a repair in the history matters only if it can be connected to the relevant function and time.
Breath testing is usually a search, but it is less intrusive than blood testing
Birchfield v. North Dakota, 579 U.S. 438, 476–77 (2016), held that a warrantless breath test may generally follow a lawful DUI arrest as a search incident to arrest, while a warrantless blood test ordinarily may not. That constitutional distinction does not dispense with arrest, advisory, implied-consent, foundation, or evidentiary requirements in a particular case.
Review when probable cause developed, when the arrest occurred, what advisory was read, what the person said and did, and which test the officer directed. If police obtained a warrant, inspect the application and execution. If the issue is refusal or noncooperation, distinguish a conscious refusal from an instrument abort, physical inability, misunderstanding, deficient sample, or operator decision to stop the attempt.
Per-se alcohol and impairment are different theories
Current N.D. Cent. Code § 39-08-01 prohibits driving or actual physical control with an alcohol concentration of at least 0.08 percent at the time of the performance of a chemical test within two hours after driving or control. Separate provisions address being under the influence. A properly admitted concentration can support the per-se theory, but timing and foundation still matter.
A result below 0.08 does not automatically end an impairment allegation, and a result above it does not establish who drove, when the driving occurred, or whether the person was in actual physical control. The defense should force the prosecution to identify the statutory theory and connect the test to every required fact rather than treating the printout as the whole case.
The approved method supplies a testable sequence, not a black box
The deprivation period addresses recent alcohol in the mouth
Before the evidentiary sequence, the approved method requires the operator to ensure the subject has had nothing to eat, drink, or smoke for twenty minutes. This period reduces the risk that recent alcohol or another event in the mouth will distort a breath sample. Identify the precise start and end, the person responsible for observation, and any interruption, transport, restroom visit, conversation, or camera gap.
The method’s words matter more than a loose label such as “observation period.” Determine whether the operator could see the person’s face and mouth, whether another officer assumed responsibility, and whether the checklist was completed from memory afterward. Eating, drinking, and smoking are expressly named. Belching, regurgitation, vomiting, medication spray, mouthwash, chewing tobacco, and foreign material require fact-specific scientific analysis rather than automatic conclusions.
The North Dakota sequence uses internal checks around two subject samples
The approved sequence runs diagnostic, room air, first subject test, room air, standard gas, room air, second subject test, and final room air. The diagnostic checks instrument functions. Room-air steps look for alcohol or contamination in the instrument environment. The known gas standard checks response against an approved reference. The two subject samples provide an agreement check.
Read the sequence vertically, including timestamps, messages, three-digit readings, gas value, and any notation below the main rows. A clean-looking reported result can follow an earlier aborted sequence that is not attached to the prosecutor’s packet. Request every test initiated on that instrument around the relevant time and preserve native or complete records where available.
The instrument asks for end-expiratory breath
The method instructs the subject to blow steadily through a clean mouthpiece until the tone stops. The April 2025 addendum explains that the tone stops and a zero appears left of the decimal when minimum flow, time, volume, and level-slope requirements have been met. Continuing beyond that minimum does not, under the addendum, make the sample invalid merely because it is larger than the minimum.
That clarification must be read together with Hoistad. There, the evidence involved repeated blowing after the instrument’s “Please Blow Until Tone Stops” instruction, and the Court concluded the Department needed expert proof about a deviation that could affect collection and reliability. Video, operator instructions, instrument prompts, tone timing, and the exact physical conduct decide whether a case resembles approved continued exhalation or a materially different sequence.
The lower adequate subject result is reported
When both samples are adequate, the method reports the lower alcohol concentration. It also applies an agreement rule to the two samples. If the samples do not satisfy that rule, the record reports “Difference Too Great,” the sequence is invalid, and another deprivation period is required before a new sequence begins.
Do not average the two values or treat the higher value as the legal result. Record both values, their order, the time between them, the reported lower number, and whether the difference suggests changing breath alcohol, inconsistent sample delivery, residual mouth alcohol, or ordinary measurement variation. A qualified expert should decide whether the pattern supports any scientific conclusion.
A deficient sample is not necessarily the same as no test
The approved method recognizes a deficient sample when the person provides no sample or does not satisfy the instrument’s adequacy criteria. If one of two samples is adequate, the method may treat that single result as a valid test and report it. The test record uses an asterisk and explanatory message, so counsel must inspect the entire page rather than only the final field.
A single adequate sample removes the comparison available from two adequate samples. That may affect weight, interpretation, and the factual account of cooperation even if the method calls the result valid. Obtain video and medical evidence concerning asthma, chronic obstructive pulmonary disease, injury, anxiety, respiratory infection, dentures, language, hearing, instructions, and the operator’s opportunities to demonstrate the technique.
Signatures and contemporaneous completion matter
The method directs the operator to complete identifying entries, answer the clean-mouthpiece question, and sign copies of the test record and checklist. Compare handwriting, electronic fields, officer report, and video. Determine when the documents were completed and whether another person entered or corrected information.
A clerical omission does not always undermine scientific reliability. But a missing signature, conflicting operator number, wrong driver, wrong citation, impossible time, changed answer, or unexplained correction can affect authentication and credibility. Separate harmless data entry from facts that call into question which person, device, sequence, or sample is actually before the court.
A breath number is a defined measurement with limits
The statutory unit is grams per 210 liters of breath
North Dakota defines breath alcohol concentration as grams of alcohol per 210 liters of end-expiratory breath. That unit is not the same specimen as blood, and the instrument does not draw blood or calculate a person’s individual blood-to-breath ratio. The law gives breath its own concentration definition for comparison with the statutory limit.
A lawyer should avoid saying the device “converts breath into blood alcohol” unless an expert is carefully explaining the legal convention. Physiological variation may be relevant to some scientific opinions, but the per-se breath allegation turns on the defined breath measurement. Use the correct phrase—breath alcohol concentration—and preserve the reported units in motions and trial questions.
Deep-lung breath is sought because exhaled concentration changes during a blow
As a person exhales, the breath comes from different portions of the respiratory system and the measured alcohol can rise before a plateau. The instrument uses flow, time, volume, and slope criteria to identify an adequate sample. Coaching, interrupted exhalation, repeated starts, shallow breathing, hyperventilation, or breath holding may affect the profile and deserve expert review if documented.
These facts do not support a universal defense. The question is what occurred in this test and whether the instrument accepted a sample under the approved criteria. Video and instrument data are more persuasive than speculation. Ask the expert which observations matter, what the instrument records, and whether any claimed effect is large enough to be relevant.
Mouth alcohol is different from alcohol carried in exhaled breath
Residual alcohol in the mouth can create a short-lived concentration not representative of end-expiratory breath. The deprivation period, two-sample sequence, slope criteria, room-air checks, and invalidation messages are safeguards, but no safeguard should be assumed from a checked box when video or testimony shows a contrary event.
The 2025 addendum says “Range Exceeded” may indicate residual mouth alcohol. It makes that test invalid, requires the instrument to abort, and directs the operator to wait another twenty minutes while ensuring no eating, drinking, or smoking before repeating the test, or to obtain an alternative evidentiary test. Preserve the first record; it explains why the later sequence occurred.
Ambient alcohol and radio-frequency events leave their own clues
Room-air steps are designed to identify alcohol or contamination around the sample chamber. An alcohol-based sanitizer, cleaning product, chemical, or another person’s breath may be relevant if the room-air result or video supports it. The solution is not to assume every station contains interfering vapor, but to obtain the room, device, and sequence evidence.
The addendum also provides that an RFI detection produces “Inhibited – RFI” on the test record. That message is evidence of an inhibited or affected sequence, not a valid reported result to be ignored. Obtain radio use, video, device location, and every later attempt to determine how the operator responded.
Temperature and medical issues need qualified proof
Body and breath temperature, respiratory conditions, gastroesophageal reflux, diabetes, ketogenic states, and medications are frequently mentioned in breath-test litigation. The existence of a diagnosis does not prove a test error. Medical records, event timing, symptoms, instrument design, and accepted research must connect the condition to a plausible effect.
Retain the right expert for the question. A physician may address diagnosis and physiology; a forensic toxicologist may address alcohol pharmacokinetics; a breath-program or instrumentation expert may address the device and method. One witness should not be stretched beyond training, and counsel should not promise a scientific defense before the underlying records are reviewed.
Alcohol can still be rising when the test is taken
A breath result measures the time of the test, not directly the earlier driving time. North Dakota’s per-se provision expressly uses a qualifying chemical test within two hours, but timing still matters to an impairment narrative and to any expert opinion about the person’s earlier concentration. Recent drinking and incomplete absorption can produce a rising pattern.
Build a reliable timeline from receipts, video, witnesses, phone data, food, drink amounts, last-drink time, driving, stop, arrest, and both samples. Retrograde extrapolation requires assumptions about absorption and elimination. A single result does not reveal the exact drinking history, and an expert should state a range and limitations rather than invent precision.
Invalidated sequences and deviations must be analyzed precisely
“Difference Too Great” starts a new twenty-minute issue
When the two subject samples differ beyond the approved limit, the sequence is invalid. The operator must wait another twenty minutes, ensuring the subject has nothing to eat, drink, or smoke, before beginning the next Intoxilyzer sequence. The period runs to the first diagnostic step of the new sequence, not merely to the next subject blow.
In Hovey, the first sequence ended at 00:37 with “Difference Too Great,” and the next diagnostic began at 00:57. The test record timestamps and deputy’s testimony supported a full twenty minutes. That holding does not excuse a shorter interval; it shows how the record can prove compliance when the timing is actually established.
Instrument timestamps should be reconciled with every available clock
Test records commonly show hours and minutes, while video or dispatch systems may show seconds. Identify whether the instrument clock was accurate, whether the header time differs from the diagnostic time, and which event starts or ends the interval. Do not round in the government’s favor or the defense’s favor without evidence.
Compare the first invalidation message, diagnostic start, officer’s spoken announcements, room camera, body camera, dispatch, booking, and report. If clocks differ, document offset rather than selecting the preferred timestamp. Hovey confirms that testimony can supplement the printed minutes, making focused cross-examination essential.
A collection deviation can require expert testimony from the proponent
Hoistad explains the division between compliance and fair-administration proof. When a deviation concerns collecting or testing the sample and may affect reliability, an ordinary person may not know its effect. The Department then cannot simply invite the hearing officer to assume harmlessness; expert evidence may be necessary to carry its burden.
The decision does not make every paperwork irregularity fatal. Identify the precise instruction, conduct, scientific purpose, possible effect, and missing proof. The strongest challenge is concrete: the video shows X, the approved method requires Y, the difference concerns sample collection or testing, and no qualified evidence establishes that the reported result remained reliable.
Calibration and diagnostic messages require the response stated in the addendum
A “Calibration Check Out of Tolerance” or “Cal Check Out of Tolerance” message makes the test invalid under the addendum. The instrument aborts, and the operator is to repeat without delay or obtain an alternative evidentiary test. A failed named diagnostic likewise invalidates and aborts the test. The diagnostic group includes room air, voltage/current, RAM, checksum, clock, digital-signal-processing, analytical-stability, modem, and temperature-regulation checks.
A “Purge Fail” during room air means the chamber did not clear after a sample; that test is invalid and aborts. The operator must clear room air and repeat without delay or obtain an alternative. Each failure has a different prescribed response. Demand the aborted record and do not let a later number erase evidence about the device’s behavior minutes earlier.
Refusal, deficient sample, and machine invalidation are not synonyms
A refusal is a legal and factual conclusion about the person’s response to a lawful testing request. A deficient sample describes breath delivery that failed adequacy criteria. “Difference Too Great,” “Range Exceeded,” RFI inhibition, purge failure, calibration error, and diagnostic failure describe instrument outcomes. One event may lead an officer to allege another, but the labels must not be collapsed.
Review every instruction, warning, attempt, message, breath sound, physical effort, operator comment, and opportunity. A person may deliberately stop blowing, misunderstand, be medically unable, or supply one adequate sample. The government must prove the theory it asserts, and the defense should use the machine’s own vocabulary when exposing a mismatch.
The defense file should include the records behind the certificate
Obtain the complete test packet and every prior attempt
Request the signed Form 106-I8000, operational checklist, implied-consent advisory, arrest report, probable-cause report, citation, temporary permit, all test records, invalid or aborted sequences, deficient-sample messages, refusal paperwork, body-camera video, station video with audio, booking video, dispatch logs, and custody records. Preserve the original digital video and metadata, not only a compressed clip.
Ask for instrument test history around the event when relevant. Another subject’s protected identifying information can be redacted while device status and error messages remain visible. The goal is to determine whether the machine moved from failure to a reported test as the officer described and whether all records were disclosed.
Use the State’s certified online records, then authenticate the date-specific documents
The Crime Laboratory posts approved-device lists, certified-operator lists, methods, checklists, gas-standard records, and individual instrument histories. Save the records in effect on the test date because live pages change. Note the retrieval date and document title, and compare the certified version offered at hearing with the public source.
For the specific serial number, examine initial and annual inspections, calibration checks, repair or service entries, placement or location, gas-cylinder records, and any period when the device was disabled. A gap does not itself prove inaccuracy, but it can support a targeted discovery request or expert inquiry.
Meaningful access to exhibits is part of a fair administrative hearing
In Hoover v. Director, North Dakota Department of Transportation, 2024 ND 8, ¶¶ 1–20, the Court reversed because the Department admitted exhibits without giving the driver a meaningful opportunity to examine them. The Court’s concern was procedural fairness, not a technical objection for its own sake.
Before the NDDOT hearing, confirm receipt of every proposed exhibit in usable form, including incorporated Crime Laboratory records. If material is missing or becomes available only during the hearing, make a clear request to inspect, object, seek a continuance when appropriate, and explain how the omission prevents cross-examination or foundation review.
An expert should answer a defined question
Possible assignments include whether the sequence followed the effective method, whether an operator instruction changed collection, whether a medical condition could explain sample difficulty, whether an error message was handled correctly, whether the sample pair is consistent, or whether a timing opinion is scientifically supportable. Send the expert the complete record rather than a favorable excerpt.
Ask for the factual assumptions, source materials, accepted principles, limitations, and opinions in plain language. An expert should not give a legal conclusion about guilt or fair administration. The useful testimony explains what a fact means scientifically and whether a departure could affect accuracy or reliability.
The NDDOT license case begins immediately after a qualifying breath result
The temporary permit is not the hearing request
When a qualifying breath result is obtained, § 39-20-03.1 generally provides for seizure of a resident’s license and issuance of a temporary operator’s permit. The permit commonly extends driving privileges for twenty-five days or until an earlier hearing decision. It does not itself request a hearing and does not give twenty-five days to make that request.
Under § 39-20-05, the written hearing request generally must be made within ten days after issuance of the temporary permit. Use an authorized NDDOT method, verify the correct identifying information, and retain submission and delivery proof. Missing the deadline can forfeit the administrative hearing even while the criminal case remains pending.
The hearing has a defined statutory scope
In a breath-result case, the hearing generally addresses reasonable grounds, arrest, compliance with testing requirements, and whether the test showed the statutory concentration. Certified records may supply prima facie foundation. The Department carries its burden by a preponderance of the evidence, and the hearing officer applies administrative evidence rules and chapter 39-20.
Prepare subpoenas, exhibits, objections, and cross-examination before the hearing. A later criminal discovery schedule does not extend the NDDOT deadline. The administrative transcript may preserve useful testimony, but the defense should not reveal privileged strategy or assume every criminal issue is within the agency’s limited scope.
The criminal case and license case can reach different results
The prosecutor must prove the criminal charge beyond a reasonable doubt. Criminal litigation may include constitutional suppression, confrontation, discovery, expert, and jury issues beyond the administrative hearing. The license case concerns the Department’s statutory authority and usually moves much faster.
A favorable ruling in one forum does not automatically terminate the other. Calendar both cases, make objections in each, and preserve each record for its own review procedure. The short administrative deadlines make early counsel especially important in a breath-test case.
An additional test is a statutory protection
Section 39-20-02 allows the tested person to have a medically qualified person of the person’s choosing administer an additional chemical test at personal expense. The failure or inability to obtain the additional test does not preclude admission of the officer-directed test, but interference with a reasonable request may raise separate issues.
Document when the request was made, what the officer said, available facilities, custody, transportation, money, telephone access, and the time lost. An independent test is time-sensitive because alcohol concentration changes. A later private test may still assist an expert, but it does not recreate the missed opportunity at the station.
A disciplined breath-test defense follows the evidence in order
Build one synchronized timeline
Include alleged driving or control, stop, observations, arrest, advisory, transport, arrival, beginning and interruptions of the deprivation period, each diagnostic, each subject sample, every error or invalidation, permit issuance, and release. Identify the clock source for each entry and calculate offsets before claiming a contradiction.
Place the video alongside the method and checklist. Mark eating, drinking, smoking, burping, vomiting, mouth spray, restroom trips, camera gaps, operator changes, prompts, tone stops, extra blows, mouthpiece changes, and officer comments. Then compare those facts with the signed certification.
Separate legal, foundational, scientific, and factual defenses
A stop or arrest issue is not the same as an approved-method deviation. A wrong operator number is not the same as mouth alcohol. An invalid sequence is not necessarily a refusal. Rising alcohol is not a machine malfunction. State each issue under the correct rule, identify who has the burden, and request the remedy that follows.
Possible relief includes suppression, exclusion for inadequate foundation, a continuance for missing exhibits, expert limitation, an administrative reversal, a jury instruction, or fact-based reasonable doubt. Some irregularities affect weight rather than admissibility. A precise argument is more credible than saying the machine is generally unreliable.
Cross-examination should begin with the witness’s own records
Establish training, certification, recollection, method version, device serial number, start time, observation location, interruptions, instructions, messages, sample attempts, completion of forms, and whether the witness reviewed video before testifying. Use short questions tied to the printed sequence and visible events.
If expert proof is missing, identify why the claimed harmlessness is outside ordinary knowledge. If the State presents an expert, test the data actually reviewed, assumptions, method language, alternative explanations, and whether the opinion addresses this deviation rather than breath testing in general.
Trial presentation should translate the sequence without overstating it
A simple timeline can show the deprivation period, first sequence, invalidation, required interval, and second sequence. A side-by-side comparison can show the method instruction and video event. The jury should see the complete printout, including messages and sample values, rather than a cropped final number.
Concede steps that are genuinely established. Focus on the missing or contradictory proof that matters to reliability, identity, timing, or the statutory theory. Human testimony and machine records can both be mistaken; the defense earns trust by distinguishing a real issue from harmless noise.
Act before records disappear
Send preservation requests for video, native test records, device data, instrument history, station logs, communications, and physical or electronic forms. Request the NDDOT hearing within ten days. Save the permit and envelope. Identify the correct instrument and retrieve the public certified records before web pages change.
Early investigation also protects favorable evidence: a continuous video showing no refusal, a witness to recent drinking, medical records explaining breathing difficulty, or a timestamp proving the required interval was short. Delay usually helps the party relying on a summary because the underlying details become harder to reconstruct.
Frequently asked questions about North Dakota DUI breath tests
1. What breath-test machine does North Dakota use?
North Dakota’s evidentiary breath program uses approved Intoxilyzer 8000 devices. The test record should identify the North Dakota Model 8000 and a serial number that can be matched to the State Crime Laboratory’s approved register and device history.
2. Is a roadside handheld breath test the same as the station test?
No. A preliminary roadside device is used differently and may serve investigative purposes. The evidentiary Intoxilyzer sequence, approved method, operator qualification, certified records, and statutory foundation are distinct. Obtain records for every device used.
3. What happens during the twenty-minute period?
The operator must ensure the subject has had nothing to eat, drink, or smoke. Review who watched, the exact start and end, camera coverage, interruptions, and any event involving the person’s mouth.
4. Does burping automatically invalidate the result?
No automatic rule should be assumed. Document the event, method language, timing, sample pattern, instrument messages, and safeguards, then obtain qualified scientific analysis about whether it could have affected the result.
5. Why are there two breath samples?
The samples provide an agreement check and the lower adequate result is reported. If they differ beyond the approved limit, the sequence reports “Difference Too Great,” becomes invalid, and requires a new twenty-minute interval before another sequence begins.
6. When does the new twenty minutes begin after “Difference Too Great”?
It begins after the first sequence is invalidated. The operator must wait the full period before the diagnostic—the first step—of the next sequence. Hovey applied that rule to the printed times and testimony.
7. What does a deficient sample mean?
It means no breath or an inadequate breath sample was delivered under the instrument’s criteria. One adequate sample may still be reported under the method. Video, instructions, effort, health, and the entire test record matter.
8. Can asthma or another medical condition explain an incomplete sample?
Potentially, but the diagnosis must be connected to the event with medical and scientific evidence. Obtain treatment records, video, sample attempts, instrument messages, and an appropriate expert opinion.
9. What does “Range Exceeded” mean?
The 2025 addendum says it may indicate residual mouth alcohol. The test is invalid and aborts. The operator must wait another twenty minutes without eating, drinking, or smoking before repeating, or obtain an alternative evidentiary test.
10. What if the diagnostic or calibration check failed?
The addendum treats specified diagnostic failures and calibration-check-out-of-tolerance messages as invalid tests that abort. Preserve the failed record and compare the operator’s next step with the approved response.
11. Can radio interference affect the instrument?
The instrument has an RFI response. Under the addendum, detection produces an “Inhibited – RFI” notation. Obtain the record, radio and video evidence, and any later sequence rather than assuming either an effect or no effect.
12. Does a breath result prove what my concentration was while driving?
It measures at test time. The per-se statute addresses a qualifying result within two hours, while an opinion about an earlier concentration requires timing and assumptions about absorption and elimination.
13. Can I request the checklist and test record?
Yes. Section 39-20-02 requires the law-enforcement agency, on request, to provide or make available the operational checklist and test record. Counsel should also seek all invalid, aborted, and deficient-sample records and the supporting certified documents.
14. Can I get my own chemical test?
Section 39-20-02 permits an additional test by a medically qualified person chosen by the tested person at personal expense. Because alcohol changes with time, make and document the request promptly.
15. How long do I have to request an NDDOT hearing?
Generally ten days after issuance of the temporary operator’s permit. The permit’s twenty-five-day driving period is not the request deadline. Submit through an authorized method and preserve confirmation.
16. What records exist for the specific instrument?
The Attorney General posts individual device pages with inspections and gas-cylinder records. Request service, repair, calibration, placement, disablement, and test-history material needed to understand the device on the test date.
17. Does one unchecked box defeat the test?
Not necessarily. Determine what fact the box represents, what actually happened, whether the method required it, and whether any deviation could affect scientific accuracy or statutory foundation. Substance matters more than slogans.
18. What should I do first after a breath-test arrest?
Save the temporary permit, request the NDDOT hearing within ten days, preserve all video and test records, obtain the effective approved method and certified device records, and have counsel compare the sequence with the evidence before deadlines expire.